Source]:pulm:period:step, Source]:pulm:polarity – RIGOL DSG3000 Series User Manual

Page 92

Advertising
background image

RIGOL

Chapter 2

Command System

2-72

DSG3000 Programming Guide

[:SOURce]:PULM:PERiod:STEP

Syntax

[:SOURce]:PULM:PERiod:STEP <value>
[:SOURce]:PULM:PERiod:STEP?

Description

Set the step of the pulse modulation period.
Query the step of the pulse modulation period.

Parameter

Name

Type

Range

Default

<value>

Real

10ns to 10s

100us

Explanation

 When <value> is set in "Number" form, the default unit is s. Besides, <value>

can also be set in "Number + Unit" form, for example, 5000ms.

 The default unit of the return value is s.

 After setting the step of the pulse period, you can use the up/down direction

keys or knob to modify the period at the current step. At this point, you can

query or set the pulse period using the

[:SOURce]:PULM:PERiod

command.

Return Format

The query returns the step of the pulse modulation period, for example, 5.

Example

:PULM:PER:STEP 5000ms
:PULM:PER:STEP?

Related

Command

[:SOURce]:PULM:PERiod

[:SOURce]:PULM:POLarity

Syntax

[:SOURce]:PULM:POLarity NORMal|INVerse
[:SOURce]:PULM:POLarity?

Description

Set the polarity of the pulse modulation.
Query the polarity of the pulse modulation.

Parameter

Name

Type

Range

Default

NORMal|INVerse

Discrete

NORMal|INVerse

NORMal

Explanation

The parameter NORMal|INVerse can set the polarity of the current pulse modulating

signal to "Normal" or "Inverse".

Return Format

The query returns NORMAL or INVERSE.

Example

:PULM:POL INV
:PULM:POL?

Advertising