[:SOURce]:PULM:DWIDth

Syntax

[:SOURce]:PULM:DWIDth <value>
[:SOURce]:PULM:DWIDth?

Description

Set the width of #2 pulse of the double pulse modulating signal.
Query the width of #2 pulse of the double pulse modulating signal.

Parameter

Name

Type

Range

Default

<value>

Real

10ns to 170s-10ns

200us

Explanation

 When the modulation source is set to "Int" and the pulse mode is set to

"Double", you can use this command to set the width of #2 pulse. Otherwise,

this command is invalid.

 When <value> is set in "Number" form, the default unit is s. Besides, <value>

can also be set in "Number + Unit" form, for example, 5000ms.

 The default unit of the return value is s.

 After setting the width of #2 pulse, you can use the up/down direction keys or

knob to modify the width at the current step. At this point, you can query or set

the current step using the

[:SOURce]:PULM:DWIDth:STEP

command.

 The width of #2 pulse is limited by the minimum pulse width, pulse period and

delay and the three fulfill the following relations.

#2 pulse width ≥ #2 minimum pulse width

#2 pulse width ≤ pulse period – delay – 10ns

Return Format

The query returns the width of #2 pulse of the double pulse modulating signal, for

example, 5.

Example

:PULM:DWID 5000ms
:PULM:DWID?

Related

Commands

[:SOURce]:PULM
[:SOURce]:PULM:DWIDth:STEP
[:SOURce]:PULM:PERiod
[:SOURce]:PULM:SOURce
[:SOURce]:PULM:MODE