18 troubleshooting, Calibrating the internal reference frequency, Calibration procedure – Kenwood TS-990S User Manual

Page 271: Calibrating the internal reference frequency -1, Calibration procedure -1

Advertising
background image

18-1

18 TROUBLESHOOTING

Index

Contents

CALIBRATING THE INTERNAL REFERENCE

FREQUENCY

The internal reference frequency of the transceiver was

properly calibrated at the factory. However, if the reference

frequency needs to be calibrated due to any reason such

as use across the ages, the internal reference frequency

can be calibrated by receiving the standard wave signal

such as VVWH (Hawaii), VVH (Colorado), BPM (Xian), or

others.

At first, follow the procedure described below to enable

the transceiver to receive the standard wave signal. The

procedure below describes the procedure when the pitch

frequency is 700 Hz.

[CW/CW-R]

[RIT]

[VOX/SEL]

[AF](M)

[CW PITCH]

[HI/SHIFT]

[LO/WIDTH]

1 Press [CW/ CW-R] to select CW mode.
2 Press [RIT] to disable the RIT function.

The "RIT" LED turns Off.

3 Press [VOL/SEL] to enable the break-in.

• If the semi break-in is enabled, press [VOX/SEL] to disable

the semi break in.

• If the full break-in is enabled, press [VOX/SEL] to disable

the full break in.

• The "VOX" LED turns Off.

4 Rotate the AF control to select the 12 o'clock position.
5 Rotate the CW PITCH control to adjust the pitch to be

legible.

You can rotate the CW PITCH control until the pitch

frequency, displayed on the sub-scope center with the sub-

screen, reaches 700 Hz.

6 Rotate the LO/WIDTH control or the HI/SHIFT control.

You can rotate the HI/SHIFT control until the shift frequency

(SHIFT), displayed on the sub-scope with the sub-screen,

reaches a value of 0, and the LO/WIDTH control until the

bandwidth (WIDTH) reaches a value of 1000.

SUB-SCREEN

Note:

◆ Refer to "CONFIGURING THE REF I/O CONNECTOR" for the

configuration of the REF I/O connector.

{page 16-22}

CALIBRATION PROCEDURE

After the calibration is completed, follow the procedure

below to configure the internal reference frequency.

[F1] [F7]

F[ADV.]

[ESC]

[AF](M)

Tuning (M)

[MULTI/CH]

[MONITOR]

1 Receive a standard wave signal on the main band.

To receive the 10 MHz standard wave signal, rotate the

Tuning control to select exactly "10.000.00".
The 700 Hz beat sounds.

f

AF

=

f

reference

+ 700

[Hz]

x

f

display [MHz]

15.6

[MHz]

f

reference

Shifts from Reference Frequency

The received beat can be audible on the CW pitch frequency.

2 Press [ADV.] (F) from the Menu screen to open the

Advanced Menu screen.

3 Access Menu 05, "Reference Oscillator Calibration".
4 Press [SELECT] (F4) to allow editing of the parameter

box.

5 Hold down [CAL.T] (F7).

The 700 Hz sidetone for calibration is generated. The double-

beat occurs due to the difference of two frequencies by the

sidetone and the received audio crossing each other.
If the double-beat cannot be heard clearly, rotate the AF

control to adjust the received audio level or the MONITOR

control to adjust the sidetone audio level.

f

sidetone

= 700

[Hz]

8

[ppm]

‒+

(700 0.006

[Hz]

)

‒+