Code #03: receive utility parameters, Code #04: not supported, Code #05: not supported – INFICON MDC-260 Thin Film Deposition Controller User Manual

Page 152: Code #03: receive utility parameters -14, Code #04: not supported -14, Code #05: not supported -14

Advertising
background image

MDC-260 DEPOSITION CONTROLLER

COMPUTER INTERFACE

11-14

Response:

• In addition to the normal transmission receipt, the controller will respond

with:

Header

Interface

Address

Instruction

Code

Length

18 Data

Bytes

Checksum

Dec

255 254

1

2 18

(data)

(checksum)

Hex

FF FE 01

02 12

(data)

(checksum)

(data):

The 18 utility values, as shown in the utility data table above.

(checksum):

The checksum from the instruction code through the last data byte.

Code #03: Receive Utility Parameters

Description:

Instructs the controller to enter all the incoming utility parameters

into memory. Use the table from Code #02 to construct a message to set the
utility parameters to your desired settings.

Format/Example:

• To re-program the controller with custom utility parameters, transmit:

Header

Interface

Address

Instruction

Code

Length

18 Data

Bytes

Checksum

Dec

255 254

1

3 18

(data)

(checksum)

Hex

FF FE 01

03 12

(data)

(checksum)

(data):

The 18 utility values, as shown in the utility data table above.

(checksum):

The checksum from the instruction code through the last data byte.


Response:

• The controller will respond with the normal transmission receipt.

Code #04: Not Supported

Description:

Instruction Code #04 is not supported by the MDC-260. Attempting

to issue it to the controller will result in an “Invalid Instruction Code” error in the
message receipt.

Code #05: Not Supported

Description:

Instruction Code #05 is not supported by the MDC-260. Attempting

to issue it to the controller will result in an “Invalid Instruction Code” error in the
message receipt.