Code #10: Send Process

Description:

Instructs the controller to send all the process parameters for

Process #N to the host computer. A description of the process parameter list is as
follows:

Parameter Name

Length (bytes)

Dec. Pt. Position

Range

Units

Process name

12

*

Characters None

Edit Password

4

*

Characters None

View/Run Password

4

*

Characters None

Number of Layers

1

*

0...250

None

Total: 21 Data Bytes


Following this general information about Process #N, one or more messages are
transmitted describing all of the layer data for the layers included in the process
are transmitted. Because the MDC-260’s message size is limited to 249 data
bytes, it may take up to five messages to transmit all of the layer data (depending
on how many layers are included in the process). Each message will contain
between one (1) and sixty (60) layers. For example, if the process contains the
maximum number of layers (250), the controller will send the layer data in five
messages. The first four messages will have 60 layers and the last message will
have 10 layers. The format of each layer messages is as follows:

Parameter Name

Length (bytes)

Dec. Pt. Position

Range

Units

Layer #N Thickness

3

3

0...999900

Å

Layer #N Material #

1

*

1...8

None

Total: 4 Data Bytes per Layer


Format/Example:

• To instruct the controller to send the parameters for Process #5, transmit:

Header

Interface

Address

Instruction

Code

Length

Process #N
(one byte)

Checksum

Dec

255 254

1

10 1 5 239

Hex

FF FE

01

0A 01 05 EF