MDC-260 DEPOSITION CONTROLLER

COMPUTER INTERFACE 11-39

Code #29: Start Process

Description:

Instructs the controller to start Process #N starting from Layer #Y.


Format/Example:

• To instruct the controller to start Process #6 starting from Layer #30,

transmit:

Header

Interface

Address

Inst.

Code

Length

Process #N

(1-10)

Layer #Y

(1-99)

Checksum

Dec

255 254

1

29

2

6 30

Hex

FF FE

01

1D

02

06 1E


Response:

• The controller will respond with the normal transmission receipt.

Code #30: Send Run-Time Values

Description:

Instructs the controller to send the run-time value list. A description

of the run-time value list follows the example.

Format/Example:

• To instruct the controller to send the run-time value list, transmit:

Header

Interface

Address

Inst.

Code

Length

Checksum

Dec

255 254

1

30

0

225

Hex

FF FE

01

1E

00

E1


Response:

• In addition to the normal transmission receipt, the controller will respond

with:

Header

Interface

Address

Inst.

Code

Length

103 Data

Bytes

Checksum

Dec

255 254

1

30

103

(data) (checksum)

Hex

FF FE

01

1E

67

(data) (checksum)

(data):

The 103 data bytes represent the run-time value list.

(checksum):

The checksum from the instruction code through the last data byte.