MDC-260 DEPOSITION CONTROLLER

COMPUTER INTERFACE

11-16

Deposit Rate #2

2

52

1

0

...9999 0.1

Å/Sec

Deposit Rate #3

2

54

1

0

...9999 0.1

Å/Sec

Time

Setpoint

2 56 * 0...35999 Seconds

Ramp to Feed Time

2

58

*

0

...35999 Seconds

Feed

Power

2 60 1

0

...999 0.1%

Feed

Time

2 62 * 0...35999 Seconds

Ramp to Idle Time

2

64

*

0

...35999 Seconds

Idle

power

2 66 1

0

...999 0.1%

Maximum

Power

2 68 1

0...999

0.1%

Minimum

Power

2 70 1

0

...999

0.1%

Rate Deviation: Attn

2

72

1

0

...999 0.1%

Rate Deviation: Alarm

2

74

1

0

...999 0.1%

Rate Deviation: Abort

2

76

1

0

...999 0.1%

Sample Dwell Period

2

78

*

0

...35999 Seconds

Backup Tooling Factor

2

80

1

100

...4999 0.1%

Material Password

4

82

*

[4 ASCII Bytes]

ASCII

Total: 86 Data Bytes

* - Indicates decimal point position is not applicable.

Format/Example:

• To instruct the controller to send the parameters for Material #4, transmit:

Header

Interface

Address

Instruction

Code

Length

Material #N

(one byte)

Checksum

Dec

255 254

1

6 1 4

244

Hex

FF FE

01

06 1 04

F4


Response:

• In addition to the normal transmission receipt, the controller will respond

with:

Header

Interface

Address

Inst.

Code

Length

Material #N

(one byte)

86 Data Bytes

Checksum

Dec

255 254

1

6 87

4

(data) (checksum)

Hex

FF FE

01

06 57

04

(data) (checksum)


(data):

The 86 material parameter data bytes, as shown in the utility data table

above.
(checksum):

The checksum from the instruction code through the last data byte.

Advertising