Code #28: send controller status, Code #28: send controller status -36 – INFICON MDC-260 Thin Film Deposition Controller User Manual
Page 174

MDC-260 DEPOSITION CONTROLLER
COMPUTER INTERFACE
11-36
Code #28: Send Controller Status
Description:
Instructs the controller to send the controller status data list.
Format/Example:
• To instruct the controller to send the controller status data list, transmit:
Header
Interface
Address
Instruction
Code
Length
Checksum
Dec
255 254
1
28 0
228
Hex
FF FE
01
1C 00 E4
Response:
• In addition to the normal transmission receipt, the controller will respond
with:
Header
Interface
Address
Inst.
Code
Length
24 Data
Bytes
Checksum
Dec
255 254
1
28
24
(data) (checksum)
Hex
FF FE
01
1C
18
(data) (checksum)
(data):
The 24 data bytes represent the controller status data list.
(checksum):
The checksum from the instruction code through the last data byte.
A description of the controller status data list is as follows:
Parameter Name
Length
(bytes)
Values
Controller State
1
1 = Process ready
2 = Start layer
3 = Change pocket
4 = Change crystal
5 = Layer ready
6 = Soak rise
7 = Soak hold
8 = Rise to Predeposit
9 = Predeposit hold
10 = Establish rate
11 = Shutter Delay
12 = Deposit #1
13 = Rate ramp #1
14 = Deposit #2
15 = Rate ramp #2