MDC-260 DEPOSITION CONTROLLER

OPERATING THE MDC-260

5-8

Warning

Messages Type

Clear

Action

Min Rate&Max Power

Alarm

Abort

Max Rate&Min Power

Alarm

Abort

System Setup memory corrupted

Alarm

Halt

Process memory corrupted

Alarm

Halt

Material memory corrupted

Alarm

Halt

Rate Est. Error

Alarm

Halt

Crystal Failure

Alarm/Attn

NO/*

Halt

Source Fault

Alarm

Halt

Sensor Fault

Alarm

Halt

No Snsrs Enabled

Alarm

Halt

Time Power

Alarm

Time/Power

Rate Dev. Alarm

Alarm

*

Alarm Action

Alarm

Crystal Marginal

Alert/Attn

NO/*

Rate Dev. Alert

Alert

*

Max power Alert

Alert

*

Min power Alert

Alert

*

Alert Action

Alert

*

Xtal Fail Switch

Attention

Crystal Switch

Xtal Mrgn Switch

Attention

Crystal Switch

Rate Dev. Atten

Attention

*

Max power

Attention

*

Min power

Attention

*

Change source # X to (material name)
and press Start to continue.

Attention

Hold

Change sensor # X to crystal # X
and press Start to continue.

Attention

Hold

Attention Action

Attention

*

Press Start to resume process.

N/A

Start to continue.

Attention

*

Hold

* Sound will clear when the condition clears.
Table 5-1 Trouble Conditions and Warnings

5.9.1 DESCRIPTION

The messages in the above table are described below.

5.9.1.1

MIN RATE&MAX POWER

This message indicates that the output power is at the maximum power level set
by the Maximum Power parameter and the rate deviation is below the limit value
set in the Rate Dev. Alarm parameter. When this happens, the controller will go
into the Abort mode and the Alarm will sound.