Xtal mrgn switch, Rate dev. atten, Maximum power – INFICON MDC-260 Thin Film Deposition Controller User Manual

Page 89: Minimum power, Change pocket, Change crystal, Attention action, Press start to resume process, Xtal mrgn switch -11, Rate dev. atten -11

Advertising
background image

MDC-260 DEPOSITION CONTROLLER

OPERATING THE MDC-260

5-11

5.9.1.20 XTAL MRGN SWITCH

This message indicates the primary crystal is marginal and the sensor input has
been switched to the backup crystal. In addition, the Attention warning sounds.
Press any key to clear the sound.

5.9.1.21 RATE

DEV.

ATTEN

The deposition rate deviation error is greater than the value set in the Rate
Deviation Attention parameter.

5.9.1.22 MAXIMUM

POWER

The output power is being limited by the value set in the Maximum Power
parameter.

5.9.1.23 MINIMUM

POWER

The output power is at or below the minimum power set by the Minimum Power
parameter.

5.9.1.24 CHANGE

POCKET...

Prompts the operator to switch the source pocket to the correct position. The
process will be on hold until the Start key is pressed. There is no message if the
Control parameter is set to Auto (Source Setup Menu).

5.9.1.25 CHANGE

CRYSTAL...

Prompts the operator to switch the sensor to the correct crystal position. The
process will be on hold until the Start key is pressed. There is no message if the
Control parameter is set to Auto (Sensor Setup Menu).

5.9.1.26 ATTENTION

ACTION

This message indicates the Alert sound was initiated by an internal action.

5.9.1.27 PRESS START TO RESUME PROCESS

This message comes up when the controller is aborted during a process. After the
abort condition is removed, the operator may press the Start key to resume the
process. To cancel from resuming the process, press Reset key.