Code #22: send input setup, Code #22: send input setup -29 – INFICON MDC-260 Thin Film Deposition Controller User Manual

Page 167

Advertising
background image

MDC-260 DEPOSITION CONTROLLER

COMPUTER INTERFACE 11-29

Code #22: Send Input Setup

Description:

Instructs the controller to send the Input parameter list for Input #N.

A description of the Input parameter list is as follows:

Parameter Name

Length (bytes)

Allowable Range

Name

16

All ASCII Characters

Type *

2

0...65535 (LSB first)

True Level

1

0 = Low, 1 = High

Input Pin#

1

30...37

Total: 20 Data Bytes

* The Type parameter defines whether the input is undefined, user defined, or
internally defined. 0 = Undefined 65,535 = user defined. All other values
indicated internally defined inputs created by the MDC for source pocket and
sensor crystal position control. When clearing an input you must set the Type = 0
(00 hex). When defining an input you must set the type = 65535 (FF hex).
Internal types should only be set by the MDC.

Format/Example:

• To instruct the controller to send the parameter list for Input #4, transmit:

Header

Interface

Address

Instruction

Code

Length

Input #N

(1-8)

Checksum

Dec

255 254

1

22 1

4 228

Hex

FF FE

01

16 01

04 E4


Response:

• In addition to the normal transmission receipt, the controller will respond

with:

Header

Interface

Address

Inst.

Code

Length

Input #N

(1-8)

20 Data

Bytes

Checksum

Dec

255 254

1

22

21

4 (data)

(checksum)

Hex

FF FE

01

16

15

04 (data)

(checksum)

(data):

The 20 data bytes represent the configuration parameters the selected

Input.
(checksum):

The checksum from the instruction code through the last data byte.

Advertising