Code #18: send source setup, Code #18: send source setup -25 – INFICON MDC-260 Thin Film Deposition Controller User Manual
Page 163
MDC-260 DEPOSITION CONTROLLER
COMPUTER INTERFACE 11-25
Code #18: Send Source Setup
Description:
Instructs the controller to send the source setup parameter list for all
sources. A description of the source parameter list for each source is as follows:
Parameter Name
Length
(bytes)
Decimal Pt.
Position
Values
Units
Number of Pockets
1
*
1
...16 None
Shutter Relay Type
1
*
17 = Normally open,
18 = Normally closed,
19 = None
None
Control
1
*
21 = Direct,
22 = BCD,
23 = Indiv,
24 = Manual
None
Drive 1
*
25 = Up
,
26 = Down,
27 = Fast,
28 = Inline,
29 = Sngl Step,
30 = Dbl Step
None
Feedback Type
1
*
31 = Individual,
32 = BCD,
33 = Single Home,
34 = In Position,
35 = No Feedback
Pocket Indexer Delay
1
*
0
...99 Seconds
Source Voltage Range
1
*
36 = 2.5V,
37 = 5.0V,
38 = 10V
Volts
Shutter Delay
2
1
0
...99 (LSB first)
.1 sec
Total: 9 Data Bytes
There are two sources with 9 bytes per source for a total of 18 bytes. All 18 bytes
are sent in the same message starting with source #1.