Code #18: send source setup, Code #18: send source setup -25 – INFICON MDC-260 Thin Film Deposition Controller User Manual

Page 163

Advertising
background image

MDC-260 DEPOSITION CONTROLLER

COMPUTER INTERFACE 11-25

Code #18: Send Source Setup

Description:

Instructs the controller to send the source setup parameter list for all

sources. A description of the source parameter list for each source is as follows:

Parameter Name

Length

(bytes)

Decimal Pt.

Position

Values

Units

Number of Pockets

1

*

1

...16 None

Shutter Relay Type

1

*

17 = Normally open,
18 = Normally closed,
19 = None

None

Control

1

*

21 = Direct,
22 = BCD,
23 = Indiv,
24 = Manual

None

Drive 1

*

25 = Up

,

26 = Down,
27 = Fast,
28 = Inline,
29 = Sngl Step,
30 = Dbl Step

None

Feedback Type

1

*

31 = Individual,
32 = BCD,
33 = Single Home,
34 = In Position,
35 = No Feedback

Pocket Indexer Delay

1

*

0

...99 Seconds

Source Voltage Range

1

*

36 = 2.5V,
37 = 5.0V,
38 = 10V

Volts

Shutter Delay

2

1

0

...99 (LSB first)

.1 sec

Total: 9 Data Bytes


There are two sources with 9 bytes per source for a total of 18 bytes. All 18 bytes
are sent in the same message starting with source #1.

Advertising