Repair and maintenance, Handling precautions, Maintenance philosophy – INFICON MDC-260 Thin Film Deposition Controller User Manual

Page 187: Trouble shooting aids, Repair and maintenance -1, Handling, Precautions -1, Maintenance, Philosophy -1, Trouble

Advertising
background image

MDC-260 DEPOSITION CONTROLLER

REPAIR AND MAINTENANCE

12-1

12. REPAIR

AND

MAINTENANCE

12.1 HANDLING PRECAUTIONS
Integrated Circuits (I.C.’s) can be damaged by static discharge into their inputs.
This static discharge is the same phenomenon that produces the unpleasant shock
when one grabs a doorknob after walking across a carpet. The likelihood of static
buildup is proportional to the dryness of the air and can be particularly
troublesome in cold, dry climates, or hot desert climates.

CAUTION: To prevent damage to internal components by means of
electrostatic discharge (ESD), wear a grounded anti-static wrist strap.

In order to minimize the chance of discharging body charge into the I.C. inputs,
always handle circuit boards by the edge, avoiding contact with the connector
area. When moving a board from one surface or work area to another surface or
work area, always personally touch the new surface or location before laying
down or inserting the board so that you, the board, and the surface, or equipment,
are all at the same potential. It is wise in dry climates to minimize the amount of
movement when handling or replacing I.C.’s in circuit boards. When handing a
circuit board or I.C. to another person, always touch the person first.

Wood or paper surfaces are the most forgiving surfaces to work on. Plastic
should be avoided. Metal is O.K. as long as the metal is always touched with the
hand prior to laying down the I.C.’s or circuit boards.

P.C. boards or I.C.’s should never be placed in plastic bags unless they are of the
conductive plastic type intended for this use. These bags are typically black or
pink and are normally labeled as conductive or anti-static. If no conductive
plastic bags are available, boards or I.C.’s can be wrapped in paper, and then
placed in plastic bags or shipping bags.

If the above precautions are observed, the chance of damage will be minimal and
no problems should be encountered.

12.2 MAINTENANCE PHILOSOPHY
The MDC-260 was designed around a maintenance-free philosophy. There are no
serviceable components inside the unit. Field repair at the component level is not
recommended and indeed can void the warranty. The following sections are
intended primarily as an aid in understanding the operation of the MDC-260 and
to help in isolating problems external to the unit.

12.3 TROUBLE SHOOTING AIDS
The following table describes possible problems that could occur when
interfacing the controller with a vacuum system. With each symptom is a list of
probable causes and suggested remedy.