MDC-260 DEPOSITION CONTROLLER

COMPUTER INTERFACE

11-26

Format/Example:

• To instruct the controller to send the source setup parameter list, transmit:

Header

Interface

Address

Instruction

Code

Length

Checksum

Dec

255 254

1

18 0

239

Hex

FF FE

01

12 00

EF


Response:

• In addition to the normal transmission receipt, the controller will respond

with:

Header

Interface

Address

Inst.

Code

Length

18 Data Bytes

Checksum

Dec

255 254

1

18 18

(data) (checksum)

Hex

FF FE

01

12 12

(data) (checksum)

(data):

The 18 data bytes represent the configuration parameters of both sources.

(checksum):

The checksum from the instruction code through the last data byte.

Code #19: Receive Source Setup

Description:

Instructs the controller to enter the incoming source setup

parameters into memory. Use the table from Code #18 to construct a message to
set the source parameters to your desired settings.

Format/Example:

• To re-program the controller with custom source parameters, transmit:

Header

Interface

Address

Instruction

Code

Length

18 Data

Bytes

Checksum

Dec

255 254

1

19 18

(data)

(checksum)

Hex

FF FE 01

13 12

(data)

(checksum)

(data):

The 18 utility values, as shown in the utility data table above.

(checksum):

The checksum from the instruction code through the last data byte.


Response:

• The controller will respond with the normal transmission receipt.