Symptom Possible

Cause Remedy

a) Line voltage selection wheel
is not installed to agree with line
voltage being used.

a) Change line voltage

selection. Refer to Section
8.3.3

b) Incorrect fuse size or type.

b) Replace with correct fuse

size (or type). See Section
8.3.3

c) Shorted rectifiers in power

supply area.

c) Return unit to manufacturer

for repair.

Unit blows line fuse.

d) Shorted transformer or filter

capacitor.

d) Return unit to manufacturer

for repair.

a) Blown fuse.

a) Replace with correct fuse

size (or type). See Section
8.3.3

b) Loose connections between

LCD and main board.

b) Check and secure LCD

backlight connector to
main board J6. Check and
secure LCD data lines
cable to main board J12.

Front Panel display never
illuminates.

c) Faulty clock generator (High

Speed Clock)

c) Return unit to manufacturer

for repair.

a) Contrast level is set too low.

a) Adjust contrast (Sec 4.1.6)

LCD display shows a blank
blue screen.

b) Bad LCD display.

b) Return unit to manufacturer

for repair.

a) Contrast level is set too high.

a) Adjust contrast (Sec 4.1.6)

LCD display shows a bright
“white” screen.

b) Bad LCD display.

b) Return unit to manufacturer

for repair.

a) Defective cable or cables.

a) Replace cable(s)

b) Defective or overloaded

sensor crystal.

b) Replace crystal with a new

one.

c) Oscillator unit connected in

the wrong direction.

c) Reverse the connections.

d) Bad Oscillator unit.

d) Replace with a new

oscillator.

“Crystal Failure” message
flashes with selected sensor
properly connected.

e) Improper cable length

e)

No control voltage while
monitoring output of
selected Source.

a) Improperly wired cable, or

shorted wires.

a) Verify the cable connection

against Figure 13-1. Also,
check for electrical shorts.

a) Defective membrane keys b)

Contact

manufacturer.

Front panel control keys
non- functional.

b) Loose connections between

the membrane keys and the
main board.

b) Check and secure

membrane connector to
main board J10.