Code #31: initiate automatic data logging, Code #31: initiate automatic data logging -43 – INFICON MDC-260 Thin Film Deposition Controller User Manual

Page 181

Advertising
background image

MDC-260 DEPOSITION CONTROLLER

COMPUTER INTERFACE 11-43

Code #31: Initiate Automatic Data Logging

Description:

This operation allows the computer to setup the MDC-260 to

automatically output selected run-time values to the computer port every 100
milliseconds. The values to be sent are determined when this instruction is
issued. Each value can be selected to be sent by setting its corresponding bit = 1.
All values with a bit setting = 0 will not be sent.

The run-time values available to be sent are the same as in Code #30. The
difference is that this instruction allows you to select which values to send and
once selected, the values are sent automatically until the next “Initiate Automatic
Data Logging” instruction or the controller power is removed.

Format/Example:

• To instruct the controller to initiate automatic data logging, transmit:

Header

Interface

Address

Inst.

Code

Length

5 Data Bytes

Checksum

Dec

255 254

1

31

5

(data) (checksum)

Hex

FF FE

01

1F

05

(data) (checksum)

(data):

The 5 data bytes represent the selection bits determining what values to

send. (checksum): The checksum from the instruction code through the last data
byte.

• To instruct the controller to stop sending data, transmit:

Header

Interface

Address

Inst.

Code

Length

5 Data Bytes

Checksum

Dec

255 254

1

31

5

0, 0, 0, 0, 0

Hex

FF FE

01

1F

05

00, 00, 00,