Instruction descriptions, Code #00: remote keypress activation of controller, Instruction – INFICON MDC-260 Thin Film Deposition Controller User Manual

Page 149: Descriptions -11

Advertising
background image

MDC-260 DEPOSITION CONTROLLER

COMPUTER INTERFACE 11-11

11.8 INSTRUCTION DESCRIPTIONS
The following section describes all of the valid instructions and shows examples
of how they are used. Each example is show in decimal [Dec] and hexadecimal
[Hex]. Each successful command will immediately prompt the MDC-260 to send
a transmission receipt, as described previously. If the message asked the MDC-
260 to transmit data back to the host computer, an example of that data is also
shown. Assumptions about the status of the MDC-260’s parameter data are
explicitly stated.

In parameter data tables, entries in bold represent the factory default values. In
the instruction examples, entries in bold represent variable fields that may change
depending on the data transmitted. If a checksum is provided in the example, it
will need to be re-calculated if any of the variable data changes.

Note: All examples assume the target controller’s USB Interface Address is 01.

Code #00: Remote Keypress Activation of Controller

Description:

Simulates a physical keypress of the MDC-260’s front panel

keypad. The valid key codes are shown in the following table:

Remote Activation Code

Description

1

START key

2

ABORT key

4

RESET key

8

SHUTTER key

16

MANUAL key

32

PROGRAM key

64

GRAPH key

128

ZERO key


Format/Example:

• To simulate a “Zero” keypress (Activation Code = 8), transmit:

Header

Interface

Address

Instruction

Code

Length

Key Code

(Data)

Checksum

Dec

255 254

1

0 1

8 246

Hex

FF FE 01

00 01

08 F6

Response:

• The controller will respond with the normal transmission receipt.