Code #20: send sensor setup, Code #20: send sensor setup -27 – INFICON MDC-260 Thin Film Deposition Controller User Manual
Page 165

MDC-260 DEPOSITION CONTROLLER
COMPUTER INTERFACE 11-27
Code #20: Send Sensor Setup
Description:
Instructs the controller to send the sensor setup parameter list for all
sensors. A description of the sensor parameter list is as follows:
Parameter Name
Length
(bytes)
Decimal Pt.
Position
Values
Units
Number of crystals
1
*
1
...16 None
Shutter Relay Type
1
*
17 = Normally open,
18 = Normally closed,
19 = None
,
20 = Dual
None
Control
1
*
21 = Direct,
22 = BCD,
23 = Indiv,
24 = Manual
None
Drive 1
*
25 = Up
,
26 = Down,
27 = Fast,
28 = Inline,
29 = Sngl Step,
30 = Dbl Step
None
Feedback Type
1
*
31 = Individual,
32 = BCD,
33 = Single Home,
34 = In Position,
35 = No Feedback
None
Pocket Indexer Delay
1
*
0
...99 Seconds
Total: 6 Data Bytes
There are two sensors with 6 bytes per sensor for a total of 12 bytes. All 12 bytes
are sent in the same message starting with sensor #1.
Format/Example:
• To instruct the controller to send the sensor setup parameter list, transmit:
Header
Interface
Address
Instruction
Code
Length
Checksum
Dec
255 254
1
20 0
235
Hex
FF FE
01
14 00
EB