MDC-260 DEPOSITION CONTROLLER
COMPUTER INTERFACE 11-31
Format/Example:
• To instruct the controller to send the parameter list for Output #2,
transmit:
•
Header
Interface
Address
Instruction
Code
Length
Output #N
(1-8)
Checksum
Dec
255 254
1
24 1 2 228
Hex
FF FE
01
18 01 02 E4
Response:
• In addition to the normal transmission receipt, the controller will respond
with:
Header
Interface
Address
Inst.
Code
Length
Output #N
(1-8)
43 Data
Bytes
Checksum
Dec
255 254
1
22
44
2 (data)
(checksum)
Hex
FF FE
01
16
2C
02 (data)
(checksum)
(data):
The 43 data bytes represent the configuration parameters the selected
Output.
(checksum):
The checksum from the instruction code through the last data byte.