39. Backup Crystal # (1 to 16)

This parameter is available when the Crystal Fail parameter (37 above) is set to
Switch. This parameter defines the backup crystal number. Note that for a dual-
crystal sensor head, this parameter should be set to 1. For a multiple-crystal
sensor head, set it to 2.

40. Backup Tooling Factor (10.0% to 499.9%)

This parameter defines the tooling factor for the backup sensor head. Note that
for a multiple-crystal sensor head, this parameter should be set to the same value
as the [primary] Tooling Factor parameter (8, above).

41. Material Password (4 digit string)

This parameter defines the edit password for the material. If the password is set
to anything other than 0000 it will not be displayed, and when you attempt to edit
the material you will first be asked to enter the correct password.