7.3

Specification

Type:

Noise Dosimeter according to ANSI S1.25-1978 Class 2A & 2C*,
Personal Sound Exposure Meter according to IEC 1252:
1993 and BS 6402: 1994,
Type 2 Sound Level Meter according to IEC 651 and
ANSI S1.4 1983,
Type 2 Integrating Sound Level Meter according to
IEC 804 and BS 6402: 1983.

Note*:

See also User adjustable parameters and Table 1 (Section 2.2).

Detector:

True RMS:

Pulse range: >63 dB,
200 Pa.

Peak:

143 dB.

Sound level range primary indication: (at 1kHz)

70 - 140 dB(A) and dB(C),
70 - 133 dB(A) and dB(C) measured according to IEC 651,
70 - 130 dB(A) and dB(C) measured according to ANSI S1.25.

The indicated range is subject to the effects of A- or
C-weighting at other frequencies.

Nominal dynamic range: 70 dB.

Noise dose range:

0.00001 - 2 x 10

9

(shown as 2000E6) % Dose.

Sound exposure range:

0.00001 - 640 000 Pa

2

h.