Casella CEL CEL-281 User Manual

Page 132

Advertising
background image

CEL-6225 Dust Cover has no measurable effect on acoustic
performance of microphone.

Electrical substitution of microphone input for
calibration:

Not recommended to be performed by unqualified personnel.

The procedure requires the microphone assembly to be replaced
by the CEL-6100 Dummy Microphone and recommends use of
an oscillator with 600

Ω impedance.

Consult CEL for precise details.

Operating environment:

Temperature:

-10 to +50

o

C,

(Effect on

accuracy:

±0.5 dB re 20

o

C at 65% RH),

Atmospheric

Pressure:

101.3 kPa

±10%

(1013 mbar, 760 mmHg),

Humidity:

30 to 90% RH (non-condensing),

(Effect on

accuracy:

±0.5 dB re 65% RH at 40

o

C),

Magnetic Fields: Negligible effect,
Vibration:

Negligible effect.

Storage environment:

(without batteries)

Temperature:

-10 to +50

o

C,

Humidity:

30 to 95% RH (non-condensing).

Consult the battery Manufacturer for storage conditions with
batteries.

Technical Information

CEL-281 Operator's Handbook - Page 123

Advertising