If the logging run is still in progress, the display is dynamic and may be seen to

increment to new values.

Note

:

The time constant used for the L

)8/

value displayed is the constant set at the

time the data was logged.

While average noise levels are being displayed, pressing the

key allows the results to be obtained from four different

exchange rate values (Q). The value in current use will be
shown against the Q legend on the display. (Remember that
L

eq

is the L

AVG

value calculated with Q = 3.)

While average noise levels are being displayed, pressing the

key selects the next from a series of up to five threshold

levels for use in the calculation. The newly selected threshold

level will be shown on the display. Then pressing the

key

again returns to the average noise level that was being in-
spected, but now shows it calculated with the new threshold
level.

All five threshold levels should be available, but if the criterion
level has been set to a value lower than 90 dB, only threshold
levels

≤ the criterion level will be available as described in Sec-

tions 2.2.2 and 2.2.3. Therefore when a threshold level is re-
quired that is higher than those currently available, the criterion
level must be increased even though it has no part in the par-
ticular L

AVG

calculation.

While Fast or Slow maxima are being displayed, pressing the

key displays the time at which the maximum level oc-

curred. Pressing

again shows the date at which the maxi-

mum occurred. Then pressing the

key twice (or once)

returns the display to the previously indicated maximum level.

How It Works