INFICON STC-2002 Thin Film Deposition Controller Operating Manual User Manual

Page 116

Advertising
background image

p STC-2002

DEPOSITION CONTROLLER

y

SECTION 3.XX

e page 114 of 276 ^

SECTION 3.19

OPTions/INFo: product identification, diagnostics, etc.




OPT

IONS

/INF

O

: From the Run Time screen, press the fixed MENU key. Press the EXECUTIVE MENU

key on the MAIN menu. (Press the BYE key to return to the MAIN menu or press the fixed STATUS key
to return to the Run Time screen.) OPT/INF is a selection on the EXECUTIVE menu. The information is
intended to be relayed to the factory for the purpose of identification and diagnostics. The information
strings contain: software Version #s, checksums, serial#s, build codes, installed option cards, option codes,
error codes, PCB hex code identifiers, frequency. There can be up to six lines of info per page, MOR

E

shows next page.

Main

Menu

Executive

Menu



Options/Info, page 1 of 6 (user interface bd.)

Options/Info, page 2 of 6 (source sensor bd.)

S/W Version, Build/CRC, SCUP COV

S/W code version / CRC each bd.

Serial number / Build type

OK

=

present, enabled, working

. OK-Unused = not enabled

Bootloader Build: CRC of BTLR code

MIA! = enabled but not physically present

UBAS min. ver of UIFC/SBAS min. ver of SCUP

Incompat

ible

! = beyond s/w family compatibility

intercommunications

Failed

=

source

sensor

board

has

failed

Error Code (Sys Trap)

BHIE: if not zero, cycle power off/on, (interrupt

[power

up

only]

not

rec'd

from

1

or

more

SS

bds.

[hold-bus]

)



Options/Info, page 3 of 6 (input/output bds.)

Options/Info, page 4 of 6

Empty = no module detected, config'd empty

SN / Build type

In-Disabled = disabled input card in this slot

S/W version

Out-Disabled = disabled output card in this slot

Configuration X CRC

Input = input card detected in this slot

Bootloader CRC

Output = output card detected in this slot

Covenant

In-Mia = missing card, enabled as if input present

Error Code (Sys Trap)

Conflict-I = have output card where input bd called [power up only]
Out-Mia = missing card, enabled as if output present
Conflict-O = have input card where output bd called
??? = reply out of range [power up only]

Advertising