INFICON STC-2002 Thin Film Deposition Controller Operating Manual User Manual

Page 227

Advertising
background image

p STC-2002

DEPOSITION CONTROLLER

y

SECTION 6.XX

e page 225 of 276 ^

18 PCD_index Indexing to pocket

19 PCD_fhold Film Holding before rise/soaks.

______________________________________________________________________

Parameter 10

Crystal Shutter

Status

Value Range

0,1

Units Open,

Closed

This parameter indicates if the crystal shutter is closed. A returned value of 0 indicate that the shutter is
open (making measurements) and a value of 1 indicates that it is closed. This is a read only parameter.
access: XTLSHUT xtal shutter state, true(1) when using a xtal, false(0) when ignoring data from crystal,
shuttered off. Identical to XTLUSMAP except during sample/hold mode, sensors mapped off to 0 when
hiding. One per active sensor ident (1..8) LSB to MSB.

0 ALL SENSOR SHUTTERS CLOSED, SENSORS NOT USED

1 Sensor # 1 XTLSHUT OPEN

2 Sensor # 2 XTLSHUT OPEN

4 Sensor # 3 XTLSHUT OPEN

8 Sensor # 4 XTLSHUT OPEN

16 Sensor # 5 XTLSHUT OPEN

32 Sensor # 6 XTLSHUT OPEN

64 Sensor # 7 XTLSHUT OPEN

128 Sensor # 8 XTLSHUT OPEN

255 ALL SENSOR SHUTTERS OPENED

______________________________________________________________________

Parameter 11

Sample/Hold Status

Value Range

0 To 2

Units None
This parameter indicates the status of sample/hold operation according to the following table:

0

Sample/Hold Not Active

1 Sampling
2 Holding
This is a read only parameter.
access: SPLHLD sample hold status, of SPLHLD_class.

0 SPLHLD_NUL not being cycled, so current state not important.

1 SPLHLD_SAM sampling, xtal being read for control.

2 SPLHLD_HOL holding, xtal being covered, pretend rate being used to accumulate thickness.

______________________________________________________________________

Parameter 12

Max. Power Limit

Status

Value Range

0,1

Units

False / True

This parameter indicates if the STC-2002 is at the max. power limit (indicated by a flashing power value on
the RunTime display). A value of 0 returned indicates that the STC-2002 is not at the max. power limit.
This is a read only parameter.
access: MAXPL true(1)/false(0) if at max power limit now. One per active source ident (1..8) LSB to
MSB.

0 No sources at MAX power limit.

1 Source # 1 at Max Power limit.

2 Source # 2 at Max Power limit.

4 Source # 3 at Max Power limit.

8 Source # 4 at Max Power limit.

16 Source # 5 at Max Power limit.

32 Source # 6 at Max Power limit.

64 Source # 7 at Max Power limit.

128 Source # 8 at Max Power limit.

255 ALL sources at MAX power limit.