p STC-2002

DEPOSITION CONTROLLER

y

SECTION 6.XX

e page 199 of 276 ^

Parameter 1

Density

access:

DENS

Value Range

0.40 - 99.99

Units

gm/cc

Reads Or Sets The Density Parameter depending which prefix letter A or B is used.
______________________________________________________________________

Parameter 2

Z Factor

access:

ZRAT

io

Value Range

0.100 - 9.999

Units None
Reads Or Sets The Z Factor Parameter depending which prefix letter A or B is used.
______________________________________________________________________

(Obsolete)

Parameter 3

Tooling Xtal 1

access:

OBSO

LETE

Was sensor 1 tool

Value Range

10.0 - 400.0

Units

Percent

______________________________________________________________________

(Obsolete)

Parameter 4

Tooling Xtal 2

access:

OBSO

LETE

Was sensor 2 tool

Value Range

10.0 - 400.0

Units

Percent

Reads or sets depending which prefix letter A or B is used, the tooling factor parameters.
These correspond to the two sensors available on the STC-2002.
______________________________________________________________________

(Obsolete)

Parameter 5

Start Xtal #

access:

OBSO

LETE

Was initial channel

Value Range

0,1

Units

0 is Sensor 1, 1 is Sensor 2

This indicates which of the 2 sensor crystals will be used for a deposition run.

______________________________________________________________________

Parameter 6

Setpoint Thk Lim

access:

STHK

Setpoint

Thickness

Limit

Value Range

0.000 - 999.999

Units

Kilo

Angstroms

The setpoint thickness parameter is used to provide a comparison point for the STC-2002
thickness setpoint event. This event will be triggered whenever the thickness display
equals or exceeds the setpoint value.
______________________________________________________________________


(non-sequencing mode only)

Parameter 7

Final Thickness Limit (Trigger)

access:

FTHK final thickness, end of run trigger

Value Range

0.000 - 999.999

Units

Kilo

Angstroms

Used to stop the deposition process in non-sequencing mode. Non-sequencing mode does
not use a process program but rather a film, which is invoked by making it the next active
film and pressing the START key sequence to begin the deposition process. It has,
therefore, one implicit process. The Final Thickness Limit Trigger stops the selected film
deposition when the thickness reaches this user entered thickness value. This value is
analgous to the thickness value in a process program when in sequencing mode.