INFICON STC-2002 Thin Film Deposition Controller Operating Manual User Manual

Page 207

Advertising
background image

p STC-2002

DEPOSITION CONTROLLER

y

SECTION 6.XX

e page 205 of 276 ^

condition as configured. A programmed value of 0 disables this test. A value of one
gives the tightest limits (5%).

______________________________________________________________________

Parameter 39

Xtal Stability

access:

XLQU crystal oscillation quality limit

Value Range

0 - 9

Units

None

This parameter is used to set a threshold on acceptable sensing crystal performance.
When the threshold is exceeded a sensor fail condition is assumed and the system will
process this condition as configured. A programmed value of 0 disables this test. A value
of one gives the tightest limits.

______________________________________________________________________

Parameter 40

Xtal Life Bounds

access:

LIFB life bounds for I/O event

Value Range

0.0 - 100.0

Units

Percent

This parameter is used to provide a setpoint on the allowable sensing crystal usage. This
setpoint may be utilized through the I/O system as a warning or interlock that sensing
crystals need attention.

______________________________________________________________________

Parameter 41

Plot Vertical Scale

access:

VSCL vertical scale units, 1..5..10..etc.

Value Range

0-4

Units

1,5,10,50,100 Å/s Or % Power As Selected

This parameter sets the vertical scale factor of the graphics display on the RunTime
screen.

______________________________________________________________________

Parameter 42

Plot Horizontal Scale

access:

HSPL horizontal sampling, 1,2,3.. samples per point

Value Range

1-600

Units

Samples Per Display Point

This parameter sets the horizontal sweep rate of the graphics display on the RunTime
screen. Samples are taken at a rate of 4 per second. A programmed value of 1 will display
every sample. The x-axis is 200 samples wide on the RunTime display.

______________________________________________________________________

Parameter 43

Data Plot Type

access:

IPLT initial plot variable, chosen at start time

Value Range

0-2

Units

0 is Rate, 1 is Rate Deviation, 2 is Power

This parameter selects the type of information to be displayed on the RunTime graphics
screen.

______________________________________________________________________

Parameter 44

S

ource

S

ensor

MAP SELECT

access:

SMAP Source/Sensor map associated, 1..30

Value Range

1 - 30

Units

map#

This parameter selects the map that is to be associated with film that invokes it.

______________________________________________________________________

Parameter 45

POCKET SELECT

Advertising