p STC-2002

DEPOSITION CONTROLLER

y

SECTION 6.XX

e page 212 of 276 ^


Command: E
Mnemonic:

EXCT_rdata (read runtime data)

Parameters:

[1..61] corresponding to real time data types PWRLVL..XSTAB8 .

Description:

Requests a read back of real time (not fixed parameters) system data. See the list of
data and the corresponding ident code below.

Returns:

[RZCD_AOK][data_value] if proper data # (1..13) id was sent. Exact format of [data_value]

depends on the particular parameter for which the data was requested.

[RZCD_ILDV] if improper value for parameter argument.

[RZCD_ILSN] if too many arguments, wrong delimiters, etc..




Command E, Read RunTime Data Parameters
______________________________________________________________________

Parameter 1

Power Level Output

access index:

PWRLVL power level, in units of .1 percent, 0.. 1000, of active source.

Value Range

0 To 1000

Units

10 X Percent

This parameter reads the current power being supplied to the active control output. It is in percent of full
scale x 10. If the number 123 is returned, this represents 12.3%.
______________________________________________________________________

Parameter 2

Deposition Rate

access index:

RATEVL rate value, +- 0 .. 999.9 A/s, as 0 .. 9999

Value Range

0 To 9999

Units

10 X Å/s

This is the current deposition rate as is displayed on the RunTime display. If the number 123 is returned,
this represents 12.3 Å/s.
______________________________________________________________________

Parameter 3

Thickness

access index:

THKVL thickness value, angstroms, 0 .. +- 999999

Value Range

0 To 999999

Units

Angstroms

This is the current thickness value as is displayed on the RunTime display. The units that are displayed are
in KÅ, but the units sent across the computer interface are in angstroms. A number returned of 1234 means
1.234 KÅ.
______________________________________________________________________

Parameter 4

Phase Timer

access index:

PHSTMR phase timer, counts up or down

Value Range

0 To 5999

Units

Seconds

This returns the time into the current phase (such as deposit). This is similar to the number on the RunTime
display, except that the time is in seconds instead of minutes:seconds.
______________________________________________________________________

Parameter 5

Running Timer

access index:

RUNTMR running timer, counts up

Value Range

0 To 5999

Units

Seconds

This returns a general timer function that is always running during a process. It pegs at 5999 and resets
when you start a new process.