DEPOSITION CONTROLLER

y

SECTION 7.XX

e page 263 of 276 ^

Tooling Determination

Place a substrate in the normal holder location and deposit a film of approximately 5000

Angstroms using either bulk or calibrated density and Z-Factor values. Make sure that when doing this
calibration the tooling is set to 100%. Measure the substrate film thickness as in the density calibration
method and determine the correct Tooling Factor value by the following formula:



Substrate

Source

Sensor

TOOLING > 100%

TOOLING < 100%

Sensor

Substrate

Source

Figure- 7.1: Typical Tooling factors



Tooling % = 100

(Substrate Thickness)

(Displayed Thickness )