p STC-2002

DEPOSITION CONTROLLER

y

SECTION 5.XX

e page 184 of 276 ^

(O64). This rung serves no useful purpose other than to demonstrate the use of one of the four
soft-switch inputs and one of the uncommitted LED indicators. When the unit is running, pressing
the leftmost membrane key will light the LED for the duration of the depression. A slight lag
between activation and deactivation of switch and the tracking of the LED illumination is normal.
Pressing and releasing the front panel switch quickly enough will result in no LED lighting,
demonstrating the need to have inputs persist long enough to be sampled and recognized.


O9

O10

T131

O141

T96

T108

T114

001:

002:

003:

004:

005:

006:

007:

008:

009:

010:

011:

I18

Stopped - Crystal Bad

Crystal Fail

P109

S8

Relay 1

C8

I38

I35

I55

P112

Relay 2

Substrate Shutter Open

I38

Setpoint Thickness

S11

Relay 3

C11

Remote Input 1

Remote Input 2

Unconditional

Process Stop

I0

I1

I45

I46

I47

I48

STOP

Deposit

Rate Ramp

Deposit after Rate Ramp

Time / Power Recovery

OUTPUTS

INPUTS

Remote Input 3

Soft Node

I2

I141

Final Thickness

Remote Input 4

I3

Use Sensor 2

Soft Node

Crystal Verify

Relay4

Sensor 2 Selected

Crystal Verify

Trigger

Figure- 5.12: Factory Installed I/O Relay Ladder Notation.



SECTION 5.16

Modulo 100 Counters ID# 400-409


Ten Modulo 100 counters are incorporated into the STC-2002 I/O structure. Each counter is

represented by a unique ID number. The status of these counter states can be used as inputs with the I/O

Re

place inc

orrect dra

w

in

g.

Advertising