RS-232 Protocol

The

RS232 protocol parameter selects one of the two RS-232 protocols available. A 0 will select

the Sycon protocol and a 1 will select the ASCII protocol.


Optional: Addresses

<not implemented> If you use an optional communications card, you will need to assign bus

addresses.

Optional: Checksum Mode

<not implemented> If the optional communications card is installed, this parameter allows an

optional checksum character to be used with the message packets.

SECTION 6.4

Industrial Interface Option

<not implemented>


SECTION 6.5

Response Format

All commands received by the STC-2002 will return a response. The minimum response (not

including protocol dependent characters) is a single ASCII character. This character indicates two things.

First the success or failure of the command which was sent. One of six result types is returned, to

indicate success (AOK), illegal command token (Groups 1..16), illegal data value (number out of range for
command), illegal syntax (something other than required = sent after command token), inhibited operation
(changing the film number while depositing for instance) and I/O sequencing error.

Second, the returned response character reports the power on reset status of the STC-2002. Each

of the four response codes (see below) is modified after a reset until the ? command resets the power failure
flag. In this way each response message informs the host as to the occurrence of an instrument reset since
the last exchange. The reason this is done is so that in every communication exchange a power on reset can
be detected and the correct exception processing be performed via host software, without special polling
overhead.

Response character

Power Lost

No Reset

Message AOK

B

A

Illegal Command

G

F

Illegal Data Value

I

H

Illegal Syntax

K

J

Inhibited Operation

M

L

I/O Sequencing Error

O

N

Table 6.7: Response Character Table.



______________________________________________________________________

SECTION 6.6


Result Codes Returned: The first letter of the reply message sent from the STC-2002 to the host computer

is from the set of result codes detailed below. For each type of result one of two reply characters is
returned depending on the nature of the power up acknowledge flag of the instrument. The second
letter of the two for each type of result is returned after any previous power failures or resets (such
as when the unit is first powered up). Once this condition is recognized and handled (bringing the
host and the STC-2002 into synchronization) and the host sends the '?' command to the STC-2002,
then the power loss/reset is acknowledged by the host and replies from the STC-2002 are then
prefixed with the first letter of the particular response type.