p STC-2002

DEPOSITION CONTROLLER

y

SECTION 6.XX

e page 216 of 276 ^

Parameter 26

source 5 power level

access index:

POWR5 power level, in units of .1 percent, 0.. 1000, of source 5.

Value Range

0 to 1000

Units

.1

Percent

source 5 power level in units of .1 percent [100% looks like 1000]. Associated card needs to be physically
installed and enabled before a reboot.
______________________________________________________________________

Parameter 27

source 6 power level

access index:

POWR6 power level, in units of .1 percent, 0.. 1000, of source 6.

Value Range

0 to 1000

Units

.1

Percent

source 6 power level in units of .1 percent [100% looks like 1000]. Associated card needs to be physically
installed and enabled before a reboot.
______________________________________________________________________

Parameter 28

source 7 power level

access index:

POWR7 power level, in units of .1 percent, 0.. 1000, of source 7.

Value Range

0 to 1000

Units

.1

Percent

source 7 power level in units of .1 percent [100% looks like 1000]. Associated card needs to be physically
installed and enabled before a reboot.
______________________________________________________________________

Parameter 29

source 8 power level

access index:

POWR8 power level, in units of .1 percent, 0.. 1000, of source 8.

Value Range

0 to 1000

Units

.1

Percent

source 8 power level in units of .1 percent [100% looks like 1000]. Associated card needs to be physically
installed and enabled before a reboot.
______________________________________________________________________

Parameter 30

Sensor 1 thickness

access index:

THICK1 thick from sensor.

Value Range

0 To 999,999

Units

Angstroms

Thickness from sensor 1. 1 of 8 in a contiguous block. Associated card needs to be physically installed and
enabled before a reboot.
______________________________________________________________________

Parameter 31

Sensor 2 thickness

access index:

THICK2 thick from sensor.

Value Range

0 To 999,999

Units

Angstroms

Thickness from sensor 2. 2 of 8 in a contiguous block. Associated card needs to be physically installed and
enabled before a reboot.
______________________________________________________________________

Parameter 32

Sensor 3 thickness

access index:

THICK3 thick from sensor.

Value Range

0 To 999,999

Units

Angstroms

Thickness from sensor 3. 3 of 8 in a contiguous block. Associated card needs to be physically installed and
enabled before a reboot.
______________________________________________________________________