p STC-2002

DEPOSITION CONTROLLER

y

SECTION 6.XX

e page 197 of 276 ^

@16 = UIFC sys_trap codes. In nnnn,nnnn form.

Read at power up, can be refreshed for recent reading from UIFC via ^U?
command. Reset from SCUP via ^U! command. Note, the ^U? command
requests a new sys_trap code from UIFC, but requires a separate @16 to
interrogate the reading if changed. Once a sys_trap code is non-zero, it will not
change or be overwritten until cleared from the UIFC front panel or via the YH
command passthru or the ^U! command.

@17..@29 reserved for future UIFC codes.

@30 = ReadBack SCUP Memory Module status. Returns CRDSTAT_class integer.

0 CRDSTAT_uninst Uninstalled option.

1 CRDSTAT_error Unavailable or unpresent when spoken to.

2 CRDSTAT_unfmtRO Present but format not recognized.

3 CRDSTAT_unfmt Present but format not recognized.

4 CRDSTAT_wrprot Present but write-protected.

5 CRDSTAT_readwrite Totally available.

@31 = Readback MemModule ProdID & Serial# returns integer, integer where first

integer is 9/10/11 for FRAM memory module w/1/2/3 chips. If chip's status is
not above unformatted, then returns 0,0. Serial number in range of 1..1000 is one
which was formatted automatically on SCUP system. Factory registered ones
start at serial # 2001 and up. This leaves an additional area from 1001..2000 for
future special purpose.

@32 = Readback Data Compatibility code returns number,number where 1st number

is product code (such as COBRASCUP #4) and second number is a dialect/data
representation covenant code which determines whether the data saved to file is
expected to be interpreted for read/write of params/accounting/sequences/ and
RLLE steps. If chip's status is not above unfmt, then returns 0,0 . This
information pertains to the portable data on the CARD.

@33 = Readback Data Compatibility code returns number,number where 1st number

is product code (such as COBRASCUP #4) and second number is a dialect/data
representation covenant code which determines whether the data saved to file is
expected to be interpreted. This information pertains to the SCUP firmware's
desired/expected information profile. It must MATCH EXACTLY the card in
order to load/clone.

______________________________________________________________________
Initialization Commands Example
Command @
Response ASTC200/B15
Description

The version request message. Where the 'A' might be a 'B' if the instrument was reset
since the last power loss acknowledgment message.

______________________________________________________________________

Reset Power Lost Status

This is used to reset the power lost status of the STC-2002. See Section x3 for a complete description.

Command: ?
Mnemonic: EXCT_ackpl (acknowledge power loss)
Parameters: None
Description: Acknowledge Power Loss status.
Returns:

[RZCD_AOK] (an A or B)

______________________________________________________________________

Initialization Commands Example
Command ?
Response A
Description

The reset power loss status message. Always an A since the message resets the 'B' status

Advertising