p STC-2002

DEPOSITION CONTROLLER

y

SECTION 6.XX

e page 200 of 276 ^

______________________________________________________________________

Parameter 8

Setpoint Time Lim

it

access:

SPTM

Value Range

0:00 - 99:59

Units

Min:Sec

The setpoint time limit parameter is used to provide a comparison point for the STC-2002
timer setpoint event status.
______________________________________________________________________

Parameter 9

Soak 1 P

o

w

e

r Value

access:

PWR1 power condition power level 1

Value Range

0.0 - 100.0

Units

Percent

This parameter sets the deposition source power value to be used as the first level of soak
power in the pre-deposition cycle. This power level will be maintained for the duration of
the soak 1 timer. This level is typically used to out gas and condition the source material.
A value of 0.0% causes the rise/soak 1 phase to be skipped.
______________________________________________________________________


Parameter 10

P

o

w

e

r Ramp 1 Time

access:

RMP1 ramp time to power level 1

Value Range

0:00 - 99:59

Units

Min:Sec

This parameter sets the duration of the power ramp from zero power to the power level
set by the soak 1 power value parameter.
______________________________________________________________________

Parameter 11

P

o

w

e

r Soak 1 Time

access:

SKT1 soak time at power level 1

Value Range

0:00 - 99:59

Units

Min:Sec

This parameter sets the time that the source will remain at the soak 1 power level
following the completion of the ramp to this level.
______________________________________________________________________

Parameter 12

Soak 2 P

o

w

e

r Value

access:

PWR2 power condition power level 2

Value Range

0.0 - 100.0

Units

Percent

This parameter sets the deposition source power value to be used as the second level of
soak power in the pre-deposition cycle. This power level will be maintained for the
duration of the soak 2 timer. A value of 0.0% skips the rise/soak 2 phases.
______________________________________________________________________



Parameter 13

P

o

w

e

r Ramp 2 Time

access:

RMP2 ramp time to power level 2

Value Range

0:00 - 99:59

Units

Min:Sec

This parameter sets the duration of the power ramp from the soak 1 power value or the
idle power value to the value set by the soak 2 power parameter.

______________________________________________________________________

Parameter 14

P

o

w

e

r Soak 2 Time

access:

SKT2 soak time at power level 2