p STC-2002

DEPOSITION CONTROLLER

y

SECTION 3.XX

e page 95 of 276 ^

Pressing the NEXT FILM key brings forth the numeric entry screen that is used to select the next active
film. BA

c

K returns the previous screen and discards the entry, if any. ENT

er

accepts the digit[s] entered as

the next active film and returns to the previous screen.


If 14 were entered, upon return to the runtime screen, the message "Next Film: 14" would appear as shown
below.


Programming the unit for non-sequencing involves the same type of communications and I/O setups,
system configuration, and film parameter modifications.
Process programming does not exist for non-sequencing product use.
An added parameter is added to the film parameter list: Final Thick Limit.
The final thickness, no longer part of process programming, becomes a parameter.

SEQ: active film# is always selected by a process.

non-SEQ: active film# is selected in place of active process# (MAIN menu key label change is apparent
along with non-appearance of the REVIEW PROCESSES key).