p STC-2002

DEPOSITION CONTROLLER

y

SECTION 1.XX

e page 15 of 276 ^

S

ource

S

ensor

MAP SELECT

1 - 30

POCKET SELECT

0 – 63 Pockets

ETCHING MODE

OFF/ON


Process Step elements:

[Menu end point: Review Processes

Menu path: Main/Review Processes]

Review Processes

(Edit Process)

(Edit Process Steps [1-99]) Mode (Sk

i

p, St

o

p, Aut

o

,

End, Wai

t

), Film# (1-99), Thickness (0.0 - 999.999 KÅ)


Main Menu values:

[Menu end point: Main Menu

Menu path: fixed front panel MENU key/ Main Menu]

Next

Active

Process

(Select)*

1 of 9 processes [sequencing mode only]

Next

Active

Film

(Select)*

1 of 99 films [non-sequencing mode only]

[*NOTE: press touch key enclosing digit to select 1 of the 9 processes or 99 films (press digits + ENTer on invoked submenu)]

RunTime Screen parameters:

[Menu end point: RunTime screen

Menu path: programming Menu/ fixed front panel STATUS key/RunTime screen]

(Crystal Quality Indicator Select)

L/Q,

Loop

x, Qual xx

(Crystal Sample Select)

(not selected [blank]), SMPL, SMPL+Time, HOLD+Time

START key sequence

(fixed front panel)

(Note PROC

ess

X: message change after each START process key press)

2

STOP key

(fixed front panel)

Stops the running process

(non-seq mode: stops the 1 inherent process [film])

MANUAL key

(fixed front panel)

Switches to manual mode from a running process only. Toggle action.

ZERO key

(fixed front panel)

Zeroes the thickness value

MENU key

(fixed front panel)

Switches from RunTime screen to MAIN menu screen

STATUS key

(fixed front panel)

Switches to RunTime screen from any menu screen, to detailed power
and crystal info screens from RunTime screen

Zero Power

[non-active process]

zero

ch

annel

1-8, zero all [1

st

STATUS key press from RunTime]

Zero Thick

ness

1

zero

ch

annel

1-8, zero all, re-verify

[2

nd

STATUS key press from RunTime]

Force Fail

[non-active process]

fail

ch

annel

1-8, fail all, re-verify

[2

nd

STATUS key press from RunTime]

1

Zero thick

ness

does not affect film thickness value, only Source Sensor card value (use to set tooling factor, diagnose

problems, etc. )

2

If the PROC

ess

X: status message does not change when the START key is pressed, check for "STOP: INV

alid

XXX "

message or check the OPT/INF menu, page 2 for cards not installed which are enabled in software (MIA). Cycle AC power
OFF ON to re-sync.

System Configuration values:

[Menu end point: System Config

Menu path: Main/Executive menu/System Configuration]

LCD Contrast

/ Bias

LOW, MEDIUM, HIGH

Password Lock #

0 - 9999

(Process)

Run Number

0 to 9999

Recorder Function

Absolute

Rate,

Rate

Deviat

ion

, P

o

w

e

r, Thickness,

Computer

Remote, I/O C

on

tr

o

l, Off

Recorder Out Ch

a

n

ne

l

1 – 8 selects an analog output channel for use

(if not off and not used as a source)

Real

Clock Time

HH/MM

Real

Clock

Date

MM/DD/YY

Need S

ource

/S

ensor

Card 1

OFF/ON

Need S

ource

/S

ensor

Card 2

OFF/ON

Need S

ource

/S

ensor

Card 3

OFF/ON

Need S

ource

/S

ensor

Card 4

OFF/ON

I/O Slot 1 Type

UNUS

ED

(DISABLED) / INPUT / OUTPUT

I/O Slot 2 Type

UNUS

ED

(DISABLED) / INPUT / OUTPUT

I/O Slot 3 Type

UNUS

ED

(DISABLED) / INPUT / OUTPUT

I/O Slot 4 Type

UNUS

ED

(DISABLED) / INPUT / OUTPUT

Memory Module IFC

OFF/ON

Communication values:

[Menu end point: Comm. Setup

Menu path: Main/Executive menu/Communications Setup]

COM/IO Lock Code

0 - 9999

Keyboard Beep

OFF / ON

RS232 Baud Rate

300, 1200, 2400, 9600

RS232

Protocol

Sycon,

ASCII