Set active process, P stc-2002, Deposition controller y – INFICON STC-2002 Thin Film Deposition Controller Operating Manual User Manual

Page 95

Advertising
background image

p STC-2002

DEPOSITION CONTROLLER

y

SECTION 3.XX

e page 93 of 276 ^


Processes: Factory settings vs. purged settings.

The following example LCD displays show the difference between factory settings and purged settings for
PROCESS# 2. The first screen is the factory setting and has, for step 01, MODE=AUTO, FILM#=2, and
THICK(KA)=0.200. The second screen is the purged setting and has only the END line.

Typical

Factory

Setting

Purged

Setting


Factory (defaulted) Settings table for All Process Numbers:

Proc# 1

Proc#2

Proc#3

Proc# 4

Proc#5

Proc#6

Proc#7

Proc#8

Proc#9

STEP

01 01 01 01 01 01 01 01 01

MODE AUTO AUTO AUTO AUTO AUTO AUTO AUTO AUTO AUTO
FILM#

1 2 3 4 5 6 7 8 9

THICK 0.100 0.200 0.300 0.400 0.500 0.600 0.700 0.800 0.900

All factory Processes have END as STEP 02.

To restore factory settings or to purge settings, press the SERVICE key on the MAIN menu. This brings
forth the SERVICE menu as shown below.

Caution


Press the AS IS key once for PURGED, twice for FACTORY, thrice to return to AS IS.


[See section x3.21, Check sum validation. See section x7.3, battery and memory loss.]

Press the ACCEPT (/ restart) key to accept the configuration as described by all of the key labels. This will
also cause the unit to REBOOT with the new configuration in effect. Please see the section describing the
SERVICE menu as there are groups of settings other than those which are PROCESS# related (Film
parameters, I/O programs, etc.) that will be included in this configuration. Valuable process
programming could be lost.
No I/O programs are factory restored because of the danger involved. Purge
will, however, clear I/O programs. See section x5.15 (table 5.4) for I/O program entry.

Press BYE to leave the SERVICE menu without accepting anything changed.

If the password is forgotten and programming has not been saved in some
other media, call factory to eliminate passwords without purging memory.


SECTION 3.7

Set Active Process


General

Description: (See Section

moved here

2.9 and figure 2.12) To change the active

process (only allowed when the instrument is in IDLE and the LAYER value is "#" indicating an at rest
condition [observe the runtime screen]), press the fixed MENU key, which will produce the MAIN menu

Please be aware that changing the
memory contents by this means changes
more than just the process values. See
section X for a complete description.