Has ID type 11 indicating 96kbyte module.





DATA TRANSFER

From the main menu, press the service key to bring forth the service menu. On the service menu, press the
key labeled AS IS. This will eventually, with the 3

rd

and 4

th

presses, reveal the memory module transaction

related keys. M.M. UNIT takes what is stored in the memory module and transfers it to the STC-2002,
which is overwritten. UNIT M.M. takes what is in the STC-2002 and stores it into the memory module.
Please note that in both cases, the recipient of the data is overwritten. Pressing the Accept key finalizes the
data transfer directive. Pressing the BYE key leaves the service screen without any effect. If other keys are
pressed on the service screen, whatever is indicated on the key legends will become implemented along
with the memory module data transfer.



Other lines from the memory module Options Info screen

ID,SRL#: The ID identifies the amount of memory in the module.

9 = 32 kbytes
10 = 64 kbytes
11 = 96 kbytes

The STC-2002 memory usage, at this time, will not exceed 32 kbytes.

The SRL# is a randomly generated serial number when formatted in the field (between 1

and 1000). It is generated and assigned by the STC-2002 to the memory module when it is
formatted. The numbers 217 and 113 can be found as serial numbers among the above screen
depictions. Factory formatted modules are assigned numbers to match a bar-coded label placed on
the exterior of the module.


MM DATA: This is a number code that identifies the product with which this memory module has been

used, if formatted. The code "4,3" identifies the STC-2002 deposition controller.


SCUP RQ: This is the number code that the STC-2002 will write to a memory module when formatted.

[The intent is to first identify the memory module in terms of its former usage, if any, and to check whether it
was a deposition controller type (the MM DATA is compared with SCUP RQ). A memory module used with
another type of product will be reformatted as it is used with an

STC-2002

. These 2 number codes are

merely an identification aid for the user.]