Dwell command process, About the dwell command process, Dwell command parameter descriptions – MTS Series 793 Application User Manual

Page 195: Dwell command process 195

Advertising
background image

Dwell Command Process

MTS MultiPurpose TestWare®

Command Processes

195

Dwell Command Process

About the Dwell Command Process

The Dwell Command process is a single segment designed to hold your
command at a static level for a specified amount of time.

When you define a Dwell Command process you must specify the duration of
the command as a time or a frequency:

Time specifies the amount of time to run the single dwell command
segment.

Frequency specifies the frequency of a command cycle (two-segments).
Since a dwell command is a single segment, the actual dwell command hold
time is half of the frequency setting. (For example, a frequency setting of 2
Hz holds the command for 0.25 seconds.)

Dwell Command Parameter Descriptions

Command tab

Command Tab

I

TEM

D

ESCRIPTION

Time/Frequency

Specifies how fast the segment runs. Options are:

Time—specifies the time to run one segment.

Frequency—specifies the time to run a two-segment cycle.

Do Not Update
Counters

Disables the counters. When enabled, the Channel Counters do not increment
for this process and the counter totals are unaffected.

Channel

Specifies the channel.

Control Mode

Specifies the control mode or control-mode variable.

Advertising
This manual is related to the following products: