MTS MultiPurpose TestWare®

Program Control Process

Other Process Descriptions

402

Log Message As

Assigns a severity level to the message (in ascending levels of severity:
Diagnostic, Information, Warning, or Error) and makes the message

available to the MPT Specimen Log.

If the Station Manager Station Log and the Basic TestWare Test Log are

configured to accept messages from all sources, then MPT makes the message

available to them also.

Note

Unlike Station Manager and Basic TestWare, MPT does not generate
messages with a severity level of Fatal Error.

Each type of log will accept the message only if it has been configured to store
messages with the assigned severity level.

For instance, if the MPT Specimen Log is configured to accept messages with
a minimum severity level of Warning, and you assign Information as the

severity level for this message, then this message will not be recorded in the

MPT Specimen Log. In that instance the MPT Specimen Log would accept
messages having only Warning and Error severity levels.

You can specify the minimum severity level that each type of log will accept.
To do this:

For MultiPurpose TestWare, select the MPT Options Editor, then
select the Specimen Log tab.

For Station Manager, select Tools, then select Station Options, then
select the Station Log tab.

For Basic TestWare, select Test Setup, then select the Test Log tab.

Action Tab (part 2 of 3)

I

TEM

D

ESCRIPTION

Advertising
This manual is related to the following products: