MTS MultiPurpose TestWare®

Temperature Control Process

External Control Process Descriptions

394

Controller options with

your controller

Depending on the type of external controller you are programming, the
Temperature Control process may be configured differently. See the following
sections for special considerations.

Eurotherm controllers

Some controllers support a special cabling configuration for Eurotherm
temperature controllers. When a Eurotherm temperature controller is connected
to your system, special Temp Output 1 and Temp Input 1 resources must be
added to your hwi file.

When configured this way, the Temperature Control process sends temperature
commands to the Eurotherm controller through the J51 serial connector at 1 Hz,
and receives temperature feedback (for data acquisition) through the same
connector at 0.5 Hz (EI-BISYNCH protocol). In the Temperature Control
process parameters window, you can set a Control Tolerance and a Dwell
Period. If your configuration includes an additional analog input resource for an
external temperature conditioner, you can also set the Monitor Tolerance.

Other controllers with

Program with Feedback

If the programming channel is configured as Program with Feedback, the
process sends the MPT end level command to the temperature controller, and can
monitor temperature feedback to maintain a Control Tolerance and a Dwell
Period.

If your configuration includes an additional analog input resource for an external
temperature conditioner, you can also set the Monitor Tolerance.

Note

The Control Tolerance and Monitor Tolerance features require
independent feedback channels.

Note

In most cases, Program with Feedback channels are configured as
low-rate channels.

Other controllers with

Program Only

If the programming channel is configured as Program Only, the process sends
the MPT end level command to the temperature controller, which maintains the
control loop. In this scenario, you cannot set a Control Tolerance, Monitor
Tolerance
, or limits on temperature feedback.

Note

In most cases, Program Only channels are configured as low-rate
channels.