Periodic Time Event Process

Event Process Descriptions

378

Periodic Time Event Parameter Descriptions

Settings tab

You may type in any of the entry boxes or choose from the selection list. To
display the selection list of an entry box, double-click on the box, or position the
cursor in the edit field of the desired box and press the up or down arrow keys.

Each box may be empty (meaning any value), may contain a single value, or may
contain a list of values separated by commas. Each value may be a single value or
a range of values (two separate values joined by a dash). For example: 1-3, 5, 7,
8-13, 19 means 1 through 3, or 5, or 7, or 8 through 13, or 19.

While running or holding, the process checks the time (clock and calendar time)
at which it begins and then periodically at one-minute intervals.

After approximately the first minute after the test starts, the process triggers
when the current time matches the schedule criteria.

Settings Tab

I

TEM

D

ESCRIPTION

Minute

Accepts values from 0 to 59.

Hour

Contains selections displayed in 12 or 24 hour format based on the regional

settings of the operating system. Accepts value entries in either the 12 or 24

hour format.

Day of Week

Contains selections displayed according to the regional settings of the

operating system. Accepts entries in English or local language.

Day of Month

Accepts values from 1 to 31.

Month of Year

Contains selections displayed according to the regional settings of the

operating system. Accepts entries in English or local language.