Set Process Parameters in the Procedure Table

MTS MultiPurpose TestWare®

Creating a Practice Test Procedure

169

4. Define the Peak/Valley Data Acquisition process (Acquire Peak/Valley).

A. In the procedure table, click the Peak/Valley Acquisition (Acquire

Peak/Valley Data) process icon.

B. In the Peak/Valley Acquisition parameters pane, click the Signals tab.

Move all the signals you want to take data on to the Included list.
Typical selections for a test like this might include Time, 10kN
Actuator Force, and 10kN Actuator Displacement.

C. Click the Acquisition tab. Type and select the parameters needed for

the test, as follows:

Master Signal: 10kN Actuator Displacement

Sensitivity: 3.0 mm

D. Click the Destination tab. Type and select the parameters needed for

the test, as follows:

Buffer Size: 16000

Data Header: Doc Data

Destination: Specimen

Buffer Type: Circular

E.

Observe the default settings for the Output Units and General tabs.

Peak/Valley Acquisition
(Acquire Peak/Valley)