MTS Series 793 Application User Manual

Page 540

Advertising
background image

MTS MultiPurpose TestWare®

Profile Editor Controls and Displays

Model 793.11 Profile Editor

540

Channel Constants Window (part 1 of 2)

I

TEM

D

ESCRIPTION

Timing

Defines a constant time base for the execution of each end level. This may be a

rate, frequency, or a time, depending on the selection in the Channel Setup

window.

Count

Specifies how many times levels 1 and 2 are repeated before moving on to the

next profile entry.

Shape

Assigns an element shape for the entire profile. The choices are:

Square—Creates square waves that travel between end levels L1 and L2.

Ramp—Creates triangle waves that travel between end levels L1 and L2.

Sine—Creates sine waves that travel between end levels L1 and L2.

True Sine—Creates sine waves that travel between the mean of end levels L1

and L2.

Square Tapered—Creates square waves that taper up from 0% to 100%

amplitude at the beginning, travel between end levels L1 and L2, and taper
down from 100%-0% at the end.

Ramp Tapered—Creates triangle waves that taper up from 0-100% amplitude
at the beginning, travel between end levels L1 and L2, and taper down from

100%-0% at the end.

Sine Tapered—Creates sine waves that taper up from 0%-100% amplitude at

the beginning, travel between end levels L1and L2, and taper down from
100%-0% at the end.

True Sine Tapered—Creates sine waves that, starting from the mean of the
amplitude, taper up from 0%-100% amplitude at the beginning, travel between

end levels L1and L2, and taper down from 100%-0% at the end, returning to

mean.

Note

The Profile Editor’s graphical display will not display the tapered wave
shapes. To view tapered wave shapes, use the Scope in the Station
Manager application.

Level1

Defines level 1 to be a constant value.

Level2

Defines level 2 to be a constant value.

Advertising