Sweep Command Process

MTS MultiPurpose TestWare®

Command Processes

221

Cycles and segments

You can repeat cycles or segments. Two segments create one cycle. Repeating a
full cycle first goes to end level 1 and ends at end level 2. Repeating a segment
can end at end level 1 or end level 2 (depending on whether the count value is
odd or even).

Note

The end levels can be set to values greater than the full scale. The
command will attempt to reach the specified end level; however, the
feedback will become saturated before reaching the specified end level
and will cause a hydraulic interlock.

End levels

The two end levels specify the amplitude of the waveform. When two test
command activities with different end levels are sequenced, the transition
between them is accomplished at the rate and segment shape of the following
activity. This provides a smooth transition between test commands.

Example

Suppose you run a Sweep activity that completes a cyclic command at End
Level 2
(in this case, +2 cm). Next, suppose you run another Sweep activity in
which End Level 1 = –1 cm and End Level 2 = +1 cm. The transition segment
bridges the two commands from the +2 cm level to the –1 cm level at the time,
rate, or frequency of the second activity.

The conventional way to
define the duration of a
cyclic command is by
frequency or time.

End level

Start level

End level 2 = 5 kip

End level 1 = 0 kip

Frequency = 5Hz

Rate = 5 kip/1 second

Time = 0.1second