MTS MultiPurpose TestWare®

Data Limit Detector Process

Event Process Descriptions

350

Limit table tab

The Limit Table tab displays all the limits defined in this process. You can edit
any data limit value in this window. To enable a limit, select the check box next
to the signal name. To disable a limit, clear the check box next to the signal name.

Action

Specifies the action that occurs when the process completes.

None

The process ends without generating a message or performing an action.

Station Power
Off

The process removes hydraulic power from the station. The test can be

resumed after you reset interlocks and reapply hydraulic power to the station.

Interlock

The process triggers a hydraulic interlock and shuts down the hydraulic power

supply. The test can be resumed after you reset the interlock, reactivate

hydraulic power.

Program
Interlock

The process triggers a program interlock. The test can be resumed after you

reset the interlock.

Program Stop

The process stops the test. This action functions the same as clicking the Stop
button on the Station Controls panel.

Program Hold

The process suspends the test until you click the Run button on the MPT

control panel. This action functions the same as clicking the Hold button on
the Station Controls panel.

Settings Tab (part 3 of 3)

I

TEM

D

ESCRIPTION