MTS MultiPurpose TestWare®

Cyclic with ALC Process

Command Processes

244

Tracking tab

Number of Cells

Specifies the number of cells in each axis of the matrix (16, 32, or 64), which
determines the resolution of the correction (the higher the number, the greater

the resolution).

Channel

Selects the channel to which the compensation ranges apply. Compensation
ranges must be entered for each channel used by the profile.

Upper Compensation
Range

Specifies the upper range value and associated units of the compensation

matrix for the selected channel.

Lower Compensation
Range

Specifies the lower range value and associated units of the compensation

matrix for the selected channel.

Time Between Matrix
Saves

Selects the time interval between matrix saves. If you Hold the test, the timer
for saving the matrix also holds.

Note

If you want the matrix to be saved only at the end of the test, type a
time longer than the profile.

Matrix Tab (part 2 of 2)

I

TEM

D

ESCRIPTION

Tracking Tab

I

TEM

D

ESCRIPTION

Channel

Lists the channel(s) to which the ALC compensation will apply.

Missed Peak
Reference

Specifies the difference between the commanded peak and the actual peak
achieved that you want the application to detect.

Missed Peaks

Displays the total number of missed peaks (also stored in the log file). This

value is saved as part of the run-time state of the process (in the mps file), and
can be restored accordingly.

When you reset the procedure, the ALC missed peaks values displayed on the

Tracking tab are not immediately reset to zero. However, when you run the test

again the process will start counting (and displaying) from zero.