MTS MultiPurpose TestWare®

Failure Detector Process

Event Process Descriptions

384

General tab

Action

Specifies the action that occurs when the process completes.

None

The process ends without generating a message or performing an action.

Station Power
Off

The process removes hydraulic power from the station. The test can be

resumed after you reset interlocks and reapply hydraulic power to the station.

Interlock

The process triggers a hydraulic interlock and shuts down the hydraulic power
supply. The test can be resumed after you reset the interlock, reactivate

hydraulic power, and apply hydraulic power to the station.

Program
Interlock

The process triggers a program interlock. The test can be resumed after you
reset the interlock.

Program Stop

The process stops the test. This action functions the same as clicking the Stop

button on the Station Controls panel.

Program Hold

The process suspends the test until you click the Run button on the MPT

control panel. This action functions the same as clicking the Hold button on

the Station Controls panel.

Options

Tab (part 2 of 2)

I

TEM

D

ESCRIPTION

General Tab

I

TEM

D

ESCRIPTION

Name

Specifies the process name displayed on the Procedure or Group process
window.

Process Enabled

Enables the process. (Clear this box to disable the process.)

Execute Process

Specifies the number of times the process will be run.

Counter Type

Enables a process counter.

Options are None, Transient, or Fixed. Transient counters are displayed only
when the process is active. Fixed counters are displayed throughout the test.

Counter Label

Names the counter.

When you run the test, this name appears under Sequence Counters on the

MPT control panel.

Note

If this box is blank, the process name is used as the counter label.