MTS MultiPurpose TestWare®

Profile with ALC Process

Command Processes

238

Verify Loaded Matrix
Parameters

Specifies whether you want the application to compare the matrix parameters
entered on this Matrix window with the parameters currently loaded in the

controller’s real-time.

If you choose “No, Use Loaded Parameters. the application will disregard

any file saved on disk that matches the entered name and use the matrix

parameters currently loaded in the controller’s real-time.

If you choose ”Yes, Reload if No Match.” the application will compare the

matrix parameters entered in this Matrix window with the parameters currently
loaded in the controller’s real-time. If the parameters do not match, the

application will redefine the real-time matrix using the parameters currently

entered in this Matrix window. In this case the application will also compare
the entered parameters with those stored in any file in the current Specimen

directory that uses the entered name. If they match, the matrix values from the

disk file will be down-loaded into the real-time matrix. If not, the real-time
matrix values will be reset to indicate no correction values are present.

To use a saved compensation matrix for the current test, you must copy (using
Windows operating software) the desired matrix file into the current Specimen

directory before you type the desired matrix name (as it appears in the

Specimen directory) and definition parameters in this window. If the name or
other parameters do not match precisely, or if the desired matrix does not exist

in the current Specimen directory, the application will create a new real-time
matrix with the entered name.

Number of Cells

Specifies the number of cells in each axis of the matrix (16, 32, or 64), which

determines the resolution of the correction (the higher the number, the greater
the resolution).

Channel

Selects the channel to which the compensation ranges apply. Compensation

ranges must be entered for each channel used by the profile.

Upper Compensation
Range

Specifies the upper range value and associated units of the compensation

matrix for the selected channel.

Lower Compensation
Range

Specifies the lower range value and associated units of the compensation
matrix for the selected channel.

Time Between Matrix
Saves

Selects the time interval between matrix saves. If you Hold the test, the timer

for saving the matrix also holds.

Note

If you want the matrix to be saved only at the end of the test, type a
time longer than the profile.

Matrix Tab (part 2 of 2)