Command tab general tab – MTS Series 793 Application User Manual

Page 370

Advertising
background image

MTS MultiPurpose TestWare®

Peak/Valley Change Detector Process

Event Process Descriptions

370

Command tab

General tab

Command Tab

I

TEM

D

ESCRIPTION

Channel Reference

Selects the command channel associated with the feedback signal that you

want to monitor. If you select None, which is the default, the Cycle Time Out
control is disabled.

Use this feature if your procedure includes tapered waveforms. With the
command channel, MPT can identify and avoid monitoring the transitions that

occur during tapered cycles.

Cycle Time Out

Note

This feature is enabled only if you select a channel in the Channel
Reference
box.

Counts cycles from the beginning of the process, including cycles in tapers. If

the process does not detect a peak or valley during the specified number of

cycles (or segments), it issues a trigger.

If you type a zero, the process will issue a trigger immediately.

If you press Stop and then Run on the Station Controls panel, the process will

start counting over again.

Note

This feature can be used to detect a malfunction; for example, if a test
is run when the Master Span (on the Station Controls panel) is set
very low.

General Tab

I

TEM

D

ESCRIPTION

Name

Specifies the process name displayed on the Procedure or Group process

window.

Process Enabled

Enables the process. (Clear this box to disable the process.)

Execute Process

Specifies the number of times the process will be run.